Category

Materials characterization

Manufacturer and Model

Bruker - Elexsys-II E500

Specifications

X-band CW EPR:

  • Working frequency range: 9.2 – 9.9 GHz
  • Output microwave power in levelled mode: 200 mW;
  • Super-High Sensitivity CW-EPR Resonator
  • Unloaded Q-factor: > 16 000
  • Max modulation amplitude: 20 G
  • Window for optical access
  • Weak Pitch signal-to-noise ratio: 3000 : 1
  • Absolute sensitivity: 1.0 x 109 spins / 1 G line width
  • Low temperature measurements: 3.8 – 298 K operated with liquid helium flow cryostat
  • Measurements at 77 K with liquid nitrogen finger dewars
  • Automated goniometer for angular dependency measurements of single crystals

Q-Band CW EPR

  • Working frequency range: 33.85 – 34.15 GHz
  • Max output microwave power: 80 mW
  • Max sample access: Æ3 mm
  • Max modulation amplitude: 20 G (100 KHz)
  • Weak Pitch signal-to-noise ratio: 2000 : 1
  • Absolute sensitivity: 1 x 109 spins / 1 G line width
  • Low temperature measurements: 3.8 – 298 K operated with liquid helium low crysostat. For Q-band measurements whole cavity is cooled

Magnetic field Specs

  • Operating range: -1 … +17 kG
  • Integrated rapid scan coils
  • Impedance 3 Ω, water cooled
  • Rapid scan
  • Magnetic field ramp range: up to 200 G
  • Scan rate: up to 200 Hz

X-BAND CW ENDOR Unit DICE-II CW ENDOR/TRIPLE System

  • X-Band CW
  • Frequency modulation with up 100 kHz rate and
  • up to 5 MHz amplitude
  • Amplitude modulation with 50 % duty cycle
  • ENDOR Induced EPR (EIE)
  • 2D ENDOR vs. Magnetic Field