Category

Materials characterization 

Manufacturer and Model

SETARAM - LABSYS Evo

Specifications

Temperature Range:

  • Room temperature to 800 °C/1 600 °C (depending on the sensor used)
  • Isothermal temperature accuracy: +/- 1 °C
  • Programmable temperature scanning rate (heating and cooling): 0.01 to 100 °C/min-1

Furnace cooling:

  • 30 min (1150 °C to 50 °C)
  • 32 min (1 600 °C to 50 °C)

Maximum balance capacity: 20 g

DSC rod – Resolution: 0.4 μW / 10 μW dependent on sensor

3D Cp rod – Cp Accuracy < 2 %,

Analysis in vacuum < 10-1 mbar

Gas Flow Control                                                  

Optional: 3 carrier gases (MFC from 4 to 200 ml/min) and 1 auxiliary or reactive gas (MFC from 0.3 to 16 ml/min)*

Applications and capabilities

With LABSYS evo STA 1600 it is possible to analyse samples at temperatures as high as 1600 °C. It is used to understand the thermal behaviour of materials that are exposed to extreme temperatures during their production, service, or recycling. The tool is used for the determination of the sintering temperatures of metal powders, determine transition temperature of the glasses and crystallization temperature of composite in composite materials such as glass ceramics and other thermal characteristics of all sorts of organic and inorganic materials.